Major anion analyses are performed using IonPac As15 (4mm) column, and self regenerating suppressor ASRS Ultra (4mm); sample loop is 25 μL.
Anion analyzed:
Fluoride
retention time (min) 3.8
Chloride 4.5
Nitrite 6.2
Sulfate 7.2
Bromide
Nirate
12.6
15
Phosphate
20
Samples should be filtered through 0.45μm filters immediately after sampling and stored at 4 ºC. The shelf time of some anions is only 48 hours. The analysis is performed using calibration standard at 4 levels adjusted to concentration of the samples. Quality control and blank control is performed every 10
th sample. For a full SOP click here.
Perchlorate is separated using IonPac AS16 (2 mm) column and self regenerating suppressor ASRS Ultra II (2mm) with external water mode.
Eluent KOH
Isocratic
65 mM
Flow rate
0.25 mL/min
Total time
14 minutes
Anion analyzed:
Perchlorate
retention time
12.5 min
Detection is performed with conductivity detector in water matrixes. For complex matrixes (e.g. tissue) the IC is connected to a mass spectrometer (Finnigan AQA) with an electrospray ionization source. The settings of mass spectrometer are:
Negative mode
-ve
AQA mix
5 V
probe temperature
300 C
For further details on our pechlorate analyses see Dodds, et al. (2004), Anal. Bioanal. Chem. 379:881-887.